Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AE0909N | AE0909D |
Duration of the Analyzing | Month (2018/05) | CUM (2016/02/01~2024/08/31) | Month (2018/05) | CUM (2016/02/01~2024/08/31) |
Runs (N) | 5 | 49 | 5 | 49 |
Mean (U/gHb) | 14.8 | 14.9 | 5.3 | 5.2 |
SD | 0.2 | 0.6 | 0.1 | 0.2 |
CV (%) | 1.4 | 4.0 | 1.9 | 3.8 |
Target Value (U/gHb) | 14.4 | 14.4 | 5.8 | 5.8 |
Total Error (%) | 5.5 | 11.5 | 12.4 | 18.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | 4.1 | 6.0 | 2.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AE0909N ; Duration : 2016-02-01 to 2024-08-31 ; Target : 14.4 ; SD : 0.70 ( ●) | |
Lot No.: AE0909D ; Duration : 2016-02-01 to 2024-08-31 ; Target : 5.8 ; SD : 0.50 ( □) | |
|
Lot No.: AE0909N ; Duration : 2016-02-01 to 2024-08-31 ; Target : 14.4 ; SD : 0.70 | |
|
Lot No.: AE0909D ; Duration : 2016-02-01 to 2024-08-31 ; Target : 5.8 ; SD : 0.50 | |
|