Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2015/08) | CUM (2014/02/01~2020/07/31) | Month (2015/08) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 8 | 222 | 8 | 222 |
Mean (U/gHb) | 13.9 | 11.5 | 3.2 | 2.7 |
SD | 1.4 | 1.9 | 0.5 | 0.5 |
CV (%) | 10.1 | 16.5 | 15.6 | 18.5 |
Target Value (U/gHb) | 13.2 | 13.2 | 5.4 | 5.4 |
Total Error (%) | 25.4 | 45.9 | 72.0 | 87.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 1.5 | 0.4 | -1.3 | -1.6 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 13.2 ; SD : 1.90 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 5.4 ; SD : 0.30 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 13.2 ; SD : 1.90 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 5.4 ; SD : 0.30 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020