各檢驗單位 G6PD 定量檢驗 內部品管統計結果 

G6PD 缺乏症檢驗之實驗室內部品管計畫

Internal Quality Control Program for G6PD Quantitative Test


線上回報系統 (G6PD IQC MIS System) [登入/login]
English


財團法人預防醫學基金會 執行

首頁申請方法 檢驗記錄表品管統計 (總表一) 月統計 vs. 累計 (總表二)各單位統計及 SDI 分佈圖


各檢驗單位 G6PD 定量檢驗 內部品管統計結果(總表一)

選擇 LotNo:
選擇 Reagent Kit:
月統計表

統計月份:
  Control N (Lot No.: AD0326N) Control D (Lot No.: AD0326D)
UnitID Reagent Kit
(Code)
Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
Lab003 1 14.814.427 0.53.59.6204.9   5.04.827 0.12.18.220>6
Lab004 1 15.015.24 0.32.05.320>6   5.05.04 0.36.012.0203.3
Lab006 1 14.816.34 0.10.611.420>6   5.05.44 0.23.715.4203.2
Lab007 1 13.813.813 0.64.38.7204.7   4.54.513 0.24.48.9204.5
Lab008 1 14.814.810 0.53.46.8205.9   5.04.510 0.24.418.9202.3
Lab010 1 14.813.95 0.21.49.020>6   5.04.75 0.24.314.5203.3
Lab011 1 14.815.25 0.21.35.320>6   5.05.25 0.23.811.7204.2
Lab012 1 14.814.75 0.42.76.120>6   5.04.75 0.24.314.5203.3
Lab013 1 15.015.07 0.21.32.720>6   4.74.67 0.24.310.8204.2
Lab015 1 14.814.49 0.42.88.320>6   6.14.99 0.24.127.8200.1
Lab029 1 14.814.95 0.32.04.720>6   5.04.85 0.12.18.220>6
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
[TOP]



累計統計表

累計期間:自
  Control N (Lot No.: AD0326N) Control D (Lot No.: AD0326D)
UnitID Reagent Kit
(Code)
Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
Lab001 1 14.214.81 ---20-   4.65.01 ---20-
Lab003 1 14.814.5744 0.85.513.1203.3   5.04.9744 0.48.218.3202.2
Lab004 1 15.014.9154 0.32.04.720>6   5.04.9154 0.24.110.2204.4
Lab006 1 14.814.8117 0.64.18.1204.9   5.04.9117 0.36.114.2203.0
Lab007 1 13.813.7244 0.53.68.0205.4   4.54.5244 0.24.48.9204.5
Lab008 1 14.814.4320 0.64.211.0204.1   5.04.6320 0.36.521.0201.8
Lab010 1 14.814.6122 0.74.810.9203.9   5.04.9122 0.36.114.2203.0
Lab011 1 14.814.9122 0.32.04.720>6   5.05.1122 0.23.99.8204.6
Lab012 1 14.814.2113 0.53.511.1204.6   5.04.7113 0.24.314.5203.3
Lab013 1 15.015.0252 0.21.32.720>6   4.74.6252 0.24.310.8204.2
Lab014 1 14.814.8190 0.42.75.420>6   5.04.8190 0.24.212.3203.8
Lab015 1 14.814.4193 0.85.613.8203.1   6.15.0193 0.48.034.0200.2
Lab016 1 14.414.982 0.21.36.220>6   5.04.882 0.24.212.3203.8
Lab029 1 14.814.9145 0.32.04.720>6   5.04.9145 0.24.110.2204.4
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
[TOP]

Reagent Kit Reagent Code
Trinity1



首頁申請方法 檢驗記錄表品管統計 (總表一) 月統計 vs. 累計 (總表二)各單位統計及 SDI 分佈圖

若有任何疑問,歡迎來電(函)聯繫或指教
地 址:106926 臺北信維郵局第624信箱  電 話:(02)2703-6080
    e-mail:  傳 真: (02)2703-6070

      財團法人預防醫學基金會
          updated : 07 Jan, 2020

回首頁