Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2014/07) | CUM (2014/02/01~2020/07/31) | Month (2014/07) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 15 | 228 | 15 | 228 |
Mean (U/gHb) | 11.4 | 11.7 | 2.6 | 2.8 |
SD | 1.5 | 1.8 | 0.4 | 0.5 |
CV (%) | 13.2 | 15.4 | 15.4 | 17.9 |
Target Value (U/gHb) | 14.6 | 14.6 | 3.2 | 3.2 |
Total Error (%) | 48.2 | 50.6 | 49.5 | 48.2 |
TEa (%) | 36 | 36 | 36 | 36 |
σ | 1.1 | 1.0 | 1.1 | 1.3 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.6 ; SD : 1.00 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.2 ; SD : 0.20 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.6 ; SD : 1.00 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.2 ; SD : 0.20 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020