Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AA1223N | AA1223D |
Duration of the Analyzing | Month (2013/01) | CUM (2012/10/01~2017/04/30) | Month (2013/01) | CUM (2012/10/01~2017/04/30) |
Runs (N) | 4 | 14 | 4 | 14 |
Mean (U/gHb) | 10.7 | 10.9 | 3.4 | 3.2 |
SD | 1.1 | 0.7 | 0.4 | 0.3 |
CV (%) | 10.3 | 6.4 | 11.8 | 9.4 |
Target Value (U/gHb) | 11.3 | 11.3 | 3.3 | 3.3 |
Total Error (%) | 25.9 | 16.4 | 26.6 | 21.8 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 1.4 | 2.6 | 1.4 | 1.8 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 11.3 ; SD : 0.5 ( ●) | |
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 3.3 ; SD : 0.3 ( □) | |
|
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 11.3 ; SD : 0.5 | |
|
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 3.3 ; SD : 0.3 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020