Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BA1116N | BA1116D |
Duration of the Analyzing | Month (2024/03) | CUM (2022/06/01~2024/03/31) | Month (2024/03) | CUM (2022/06/01~2024/03/31) |
Runs (N) | 4 | 63 | 4 | 63 |
Mean (U/gHb) | 14.1 | 14.0 | 4.5 | 4.3 |
SD | 0.3 | 1.0 | 0.1 | 0.4 |
CV (%) | 2.1 | 7.1 | 2.2 | 9.3 |
Target Value (U/gHb) | 14.1 | 14.1 | 4.5 | 4.5 |
Total Error (%) | 4.3 | 15.0 | 4.4 | 23.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | 2.7 | >6 | 1.7 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-03-31 ; Target : 14.1 ; SD : 1.41 ( ●) | |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-03-31 ; Target : 4.5 ; SD : 0.45 ( □) | |
|
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-03-31 ; Target : 14.1 ; SD : 1.41 | |
|
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-03-31 ; Target : 4.5 ; SD : 0.45 | |
|