Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BJ0922N | BJ0922D |
Duration of the Analyzing | Month (2022/04) | CUM (2021/01/01~2024/09/30) | Month (2022/04) | CUM (2021/01/01~2024/09/30) |
Runs (N) | 0 | 20 | 0 | 20 |
Mean (U/gHb) | - | 14.4 | - | 5.2 |
SD | - | 0.5 | - | 0.2 |
CV (%) | - | 3.5 | - | 3.8 |
Target Value (U/gHb) | 13.8 | 13.8 | 4.7 | 4.7 |
Total Error (%) | - | 11.3 | - | 18.3 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 4.5 | - | 2.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BJ0922N ; Duration : 2021-01-01 to 2024-09-30 ; Target : 13.8 ; SD : 1.38 ( ●) | |
Lot No.: BJ0922D ; Duration : 2021-01-01 to 2024-09-30 ; Target : 4.7 ; SD : 0.47 ( □) | |
|
Lot No.: BJ0922N ; Duration : 2021-01-01 to 2024-09-30 ; Target : 13.8 ; SD : 1.38 | |
|
Lot No.: BJ0922D ; Duration : 2021-01-01 to 2024-09-30 ; Target : 4.7 ; SD : 0.47 | |
|