Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BJ0922N | BJ0922D |
Duration of the Analyzing | Month (2022/04) | CUM (2021/01/01~2024/09/30) | Month (2022/04) | CUM (2021/01/01~2024/09/30) |
Runs (N) | 3 | 27 | 3 | 27 |
Mean (U/gHb) | 14.6 | 14.4 | 4.3 | 4.7 |
SD | 0.4 | 0.5 | 0.0 | 0.3 |
CV (%) | 2.7 | 3.5 | 0.0 | 6.4 |
Target Value (U/gHb) | 14.4 | 14.4 | 4.6 | 4.6 |
Total Error (%) | 6.9 | 6.9 | 6.5 | 14.9 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | 5.7 | >6 | 2.8 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BJ0922N ; Duration : 2021-01-01 to 2024-09-30 ; Target : 14.4 ; SD : 1.16 ( ●) | |
Lot No.: BJ0922D ; Duration : 2021-01-01 to 2024-09-30 ; Target : 4.6 ; SD : 0.30 ( □) | |
|
Lot No.: BJ0922N ; Duration : 2021-01-01 to 2024-09-30 ; Target : 14.4 ; SD : 1.16 | |
|
Lot No.: BJ0922D ; Duration : 2021-01-01 to 2024-09-30 ; Target : 4.6 ; SD : 0.30 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020