| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2020/07) | CUM (2014/02/01~2020/07/31) | Month (2020/07) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 2 | 142 | 2 | 142 |
Mean (U/gHb) | 15.4 | 13.5 | 4.0 | 3.5 |
SD | - | 1.6 | - | 0.9 |
CV (%) | - | 11.9 | - | 25.7 |
Target Value (U/gHb) | 14.4 | 14.4 | 5.8 | 5.8 |
Total Error (%) | - | 30.0 | - | 91.1 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 1.2 | - | -0.8 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.4 ; SD : 1.30 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 5.8 ; SD : 0.40 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.4 ; SD : 1.30 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 5.8 ; SD : 0.40 | |
|