Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2017/06) | CUM (2014/02/01~2020/07/31) | Month (2017/06) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 0 | 58 | 0 | 58 |
Mean (U/gHb) | - | 11.1 | - | 2.7 |
SD | - | 1.2 | - | 0.3 |
CV (%) | - | 10.8 | - | 11.1 |
Target Value (U/gHb) | 11.6 | 11.6 | 2.7 | 2.7 |
Total Error (%) | - | 25.9 | - | 22.2 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 1.5 | - | 1.8 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 11.6 ; SD : 1.16 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 2.7 ; SD : 0.27 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 11.6 ; SD : 1.16 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 2.7 ; SD : 0.27 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020