Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2017/06) | CUM (2014/02/01~2020/07/31) | Month (2017/06) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 3 | 99 | 3 | 99 |
Mean (U/gHb) | 10.5 | 12.1 | 2.4 | 3.0 |
SD | 0.6 | 1.7 | 0.1 | 0.4 |
CV (%) | 5.7 | 14.0 | 4.2 | 13.3 |
Target Value (U/gHb) | 14.4 | 14.4 | 3.3 | 3.3 |
Total Error (%) | 38.5 | 44.1 | 35.6 | 35.8 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | -1.2 | 0.3 | -1.7 | 0.8 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.4 ; SD : 2.97 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.3 ; SD : 0.77 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.4 ; SD : 2.97 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.3 ; SD : 0.77 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020