| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2017/06) | CUM (2014/02/01~2020/07/31) | Month (2017/06) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 4 | 204 | 4 | 204 |
Mean (U/gHb) | 10.9 | 12.5 | 2.4 | 2.9 |
SD | 0.2 | 1.5 | 0.1 | 0.5 |
CV (%) | 1.8 | 12.0 | 4.2 | 17.2 |
Target Value (U/gHb) | 14.8 | 14.8 | 3.4 | 3.4 |
Total Error (%) | 30.0 | 39.5 | 37.7 | 49.2 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | -3.5 | 0.4 | -2.2 | 0.3 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.8 ; SD : 1.60 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.4 ; SD : 0.50 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 14.8 ; SD : 1.60 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 3.4 ; SD : 0.50 | |
|