Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AC1203N | AC1203D |
Duration of the Analyzing | Month (2017/06) | CUM (2014/02/01~2020/07/31) | Month (2017/06) | CUM (2014/02/01~2020/07/31) |
Runs (N) | 0 | 113 | 0 | 113 |
Mean (U/gHb) | - | 10.9 | - | 2.6 |
SD | - | 1.3 | - | 0.9 |
CV (%) | - | 11.9 | - | 34.6 |
Target Value (U/gHb) | 10.8 | 10.8 | 2.5 | 2.5 |
Total Error (%) | - | 24.8 | - | 73.2 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 1.6 | - | 0.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 10.8 ; SD : 1.00 ( ●) | |
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 2.5 ; SD : 0.40 ( □) | |
|
Lot No.: AC1203N ; Duration : 2014-02-01 to 2020-07-31 ; Target : 10.8 ; SD : 1.00 | |
|
Lot No.: AC1203D ; Duration : 2014-02-01 to 2020-07-31 ; Target : 2.5 ; SD : 0.40 | |
|