Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AA1223N | AA1223D |
Duration of the Analyzing | Month (2017/04) | CUM (2012/10/01~2017/04/30) | Month (2017/04) | CUM (2012/10/01~2017/04/30) |
Runs (N) | 0 | 51 | 0 | 51 |
Mean (U/gHb) | - | 15.5 | - | 4.6 |
SD | - | 1.0 | - | 0.4 |
CV (%) | - | 6.5 | - | 8.7 |
Target Value (U/gHb) | 15.6 | 15.6 | 4.6 | 4.6 |
Total Error (%) | - | 13.5 | - | 17.4 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 3.0 | - | 2.3 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 15.6 ; SD : 0.80 ( ●) | |
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 4.6 ; SD : 0.37 ( □) | |
|
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 15.6 ; SD : 0.80 | |
|
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 4.6 ; SD : 0.37 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020