Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | AA1223N | AA1223D |
Duration of the Analyzing | Month (2017/04) | CUM (2012/10/01~2017/04/30) | Month (2017/04) | CUM (2012/10/01~2017/04/30) |
Runs (N) | 0 | 45 | 0 | 45 |
Mean (U/gHb) | - | 15.2 | - | 4.4 |
SD | - | 0.9 | - | 0.5 |
CV (%) | - | 5.9 | - | 11.4 |
Target Value (U/gHb) | 15.4 | 15.4 | 4.3 | 4.3 |
Total Error (%) | - | 13.1 | - | 25.1 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 3.2 | - | 1.6 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 15.4 ; SD : 0.5 ( ●) | |
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 4.3 ; SD : 0.3 ( □) | |
|
Lot No.: AA1223N ; Duration : 2012-10-01 to 2017-04-30 ; Target : 15.4 ; SD : 0.5 | |
|
Lot No.: AA1223D ; Duration : 2012-10-01 to 2017-04-30 ; Target : 4.3 ; SD : 0.3 | |
|