Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | 20091127N | 20091127D |
Duration of the Analyzing | Month (2012/10) | CUM (2011/03/01~2014/07/31) | Month (2012/10) | CUM (2011/03/01~2014/07/31) |
Runs (N) | 3 | 68 | 3 | 68 |
Mean (U/gHb) | 14.0 | 14.0 | 4.6 | 4.6 |
SD | 0.1 | 0.2 | 0.1 | 0.1 |
CV (%) | 0.7 | 1.4 | 2.2 | 2.2 |
Target Value (U/gHb) | 14.0 | 14.0 | 4.6 | 4.6 |
Total Error (%) | 1.4 | 2.9 | 4.3 | 4.3 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | >6 | >6 | >6 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: 20091127N ; Duration : 2011-03-01 to 2014-07-31 ; Target : 14.0 ; SD : 0.3 ( ●) | |
Lot No.: 20091127D ; Duration : 2011-03-01 to 2014-07-31 ; Target : 4.6 ; SD : 0.3 ( □) | |
|
Lot No.: 20091127N ; Duration : 2011-03-01 to 2014-07-31 ; Target : 14.0 ; SD : 0.3 | |
|
Lot No.: 20091127D ; Duration : 2011-03-01 to 2014-07-31 ; Target : 4.6 ; SD : 0.3 | |
|