| Control N | Control D |
QC Control Lot No. | 20091127N | 20091127D |
Duration of the Analyzing | Month (2012/10) | CUM (2011/03/01~2014/07/31) | Month (2012/10) | CUM (2011/03/01~2014/07/31) |
Runs (N) | 4 | 118 | 4 | 118 |
Mean (U/gHb) | 14.0 | 13.7 | 4.5 | 4.4 |
SD | 0.2 | 0.6 | 0.3 | 0.2 |
CV (%) | 1.4 | 4.4 | 6.7 | 4.5 |
Target Value (U/gHb) | 14.0 | 14.0 | 4.6 | 4.6 |
Total Error (%) | 2.9 | 10.9 | 15.5 | 13.4 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | 4.1 | 2.7 | 3.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: 20091127N ; Duration : 2011-03-01 to 2014-07-31 ; Target : 14.0 ; SD : 0.5 ( ●) | |
Lot No.: 20091127D ; Duration : 2011-03-01 to 2014-07-31 ; Target : 4.6 ; SD : 0.3 ( □) | |
|
Lot No.: 20091127N ; Duration : 2011-03-01 to 2014-07-31 ; Target : 14.0 ; SD : 0.5 | |
|
Lot No.: 20091127D ; Duration : 2011-03-01 to 2014-07-31 ; Target : 4.6 ; SD : 0.3 | |
|